PRF parameter explorer

Move each control as you would on the generator and see what it does: to the field at the nerve, to the heat at the tip, and why it matters clinically.

Pulsed modeRF generator
DeliveredV
Tip average°C
Dose index
Duty cycle%
Peak currentmA
Current-on times

Output

Time

Safety limit

Generator reading

Advice for these settings

Dose at the nerve
0100 = well-placed standard300

Thermal safety
37 °C424550 °C

What to change first

    How the settings interact

    Every setting serves one of two goals: more electric field at the nerve (the presumed active ingredient) or less heat at the tip (the limiting factor). Most settings trade one against the other.

    Duty cycle = frequency × pulse width
    Heating demand ∝ voltage² ÷ impedance × duty cycleRelative to 45 V, 400 Ω, 2 Hz, 20 ms
    Dose ∝ field² × current-on time

    This tool assumes a well-placed tip. In practice, proximity to the nerve (a low sensory threshold) and pointing the tip end-on matter as much as any generator setting.

    Each setting in detail

    How the settings interact

    Pick any two settings to see how outcomes change as they vary together, with everything else held at your current values. Tap the map to move both sliders to that point.

    Colour shows
    White line and hatching: throttlingYour settings

    Same heat, different pulse shape

    Each row heats the tip as much as your current settings and gives the same dose index. Only the balance between voltage and pulse rate changes.

    Routes to a target dose

    Target

    Turning one dial at a time from your current settings. What each route costs in heat and field differs.